{"id":4884,"date":"2020-04-28T14:53:15","date_gmt":"2020-04-28T14:53:15","guid":{"rendered":"https:\/\/www.srpcontrol.com\/?p=4884"},"modified":"2023-09-21T18:44:59","modified_gmt":"2023-09-21T18:44:59","slug":"understanding-as-found-and-as-left-data-when-it-comes-to-calibration","status":"publish","type":"post","link":"https:\/\/www.srpcontrol.com\/understanding-as-found-and-as-left-data-when-it-comes-to-calibration\/","title":{"rendered":"Understanding As-Found and As-Left Data When It Comes to Calibration"},"content":{"rendered":"<p><span data-preserver-spaces=\"true\">Have you sent your instrument out for recalibration, gotten the certificate back, and seen &#8220;As-Found Calibration&#8221; or &#8220;As-Left Calibration&#8221; and wondered what this means? Here is a guide to understanding what these terms mean.<\/span><\/p>\n<p><strong><span data-preserver-spaces=\"true\">The Definition of Calibration<\/span><\/strong><\/p>\n<p><span data-preserver-spaces=\"true\">Calibration compares a measurement device (an unknown) against an equal or better standard.\u00a0<\/span><\/p>\n<p><span data-preserver-spaces=\"true\">A standard in a measurement is considered the reference. It is the one in the comparison taken to be the more correct. Calibration finds out how far the unknown is from the standard.<\/span><\/p>\n<p><span data-preserver-spaces=\"true\">A typical commercial calibration uses the manufacturer&#8217;s calibration procedure. It is performed with a reference standard at least four times more accurately than the instrument under test.<\/span><\/p>\n<p><strong><span data-preserver-spaces=\"true\">What As-Found Calibration Means<\/span><\/strong><\/p>\n<p><span data-preserver-spaces=\"true\">As-Found data is what a calibration lab finds when it calibrates on an as-received device, like a calibrator, gauge, or sensor, before any adjustments or repairs.<\/span><\/p>\n<p><strong><span data-preserver-spaces=\"true\">What As-Left Calibration Means<\/span><\/strong><\/p>\n<p><span data-preserver-spaces=\"true\">As-Left data is the calibration data when the instrument leaves the lab.<\/span><\/p>\n<p><strong><span data-preserver-spaces=\"true\">As-Found and As-Left Data Importance\u00a0<\/span><\/strong><\/p>\n<p><span data-preserver-spaces=\"true\">Suppose the As-Found data is outside of published specifications. In that case, the device must be adjusted to bring it within spec. For example, the data will sometimes be linearized if this process consists of zero, offset, and span adjustments. Sometimes limits narrower than the published specs are applied to account for uncertainties in the measuring system, known as guard banding. Another possibility is that the device needs repairing. No matter the situation, the calibration is re-run. The data generated during this recalibration is the as-left data.<\/span><\/p>\n<p><span data-preserver-spaces=\"true\">As-Found data is crucial because it will determine the validity of any tests run with the device in the past. In addition, as-Found data provides a history of the sensor&#8217;s drift characteristics over time. It also indicates how often the sensor should be calibrated and if it meets the manufacturer&#8217;s recommended calibration interval specification.<\/span><\/p>\n<p><span data-preserver-spaces=\"true\">As-Left data is also essential because it assures the operator that future tests using the device conform to expectations of inaccuracy.<\/span><\/p>\n<p><strong><span data-preserver-spaces=\"true\">Calibration Certificates Terms<\/span><\/strong><\/p>\n<ul>\n<li><u><span data-preserver-spaces=\"true\">As-Found Calibration<\/span><\/u><span data-preserver-spaces=\"true\">\u00a0\u2013 This shows the as-received device data points. If this data falls outside the published specs, or guardband if applicable, it will be issued as an as-found certificate.<\/span><\/li>\n<li><u><span data-preserver-spaces=\"true\">As-Left Calibration<\/span><\/u><span data-preserver-spaces=\"true\">\u00a0\u2013 If the As-Found data falls outside the published specs or guardband, adjustments to bring it within spec\/guardband limits are made, and the calibration is re-run to confirm it is now within requirements. This data is issued as an As-Left data certificate.<\/span><\/li>\n<li><u><span data-preserver-spaces=\"true\">As-Found and As-Left Calibration<\/span><\/u><span data-preserver-spaces=\"true\">\u00a0\u2013 In this case, the device was found to be within the spec\/guardband limits as received, so no adjustment is necessary.\u00a0<\/span><\/li>\n<li><span data-preserver-spaces=\"true\">Consequently, the device is not adjusted, the calibration data is not re-run, and the As-Left data is the As-Found data. Only one certificate is issued, marked as both As-Found and As-Left.<\/span><\/li>\n<\/ul>\n<p><strong><span data-preserver-spaces=\"true\">What Happens if the Instrument is Damaged or New?<\/span><\/strong><\/p>\n<p><span data-preserver-spaces=\"true\">If the device cannot perform an as-found calibration run as received if the device is damaged, then there will only be an As-Left Certificate. Repair work to the instrument may affect the calibration data, so any calibration data run once the unit is repaired may not be the data that would have been found if the instrument could have been calibrated as received. In this case, a lab will only do an As-Left calibration and supply only an An-Left certificate.<\/span><\/p>\n<p><span data-preserver-spaces=\"true\">This would also apply to new sensors and calibrators. However, there will only be as-left data as these instruments are only calibrated once they have completed testing and quality procedures in the factory, so they will not need to be adjusted by the calibration lab. So, the calibration certificate issued will be As-Left data.<\/span><\/p>\n","protected":false},"excerpt":{"rendered":"Have you sent your instrument out for recalibration, gotten the certificate back, and seen &#8220;As-Found Calibration&#8221; or &#8220;As-Left Calibration&#8221; and wondered what this means? Here is a guide to understanding what these terms mean. The Definition of Calibration Calibration compares a measurement device (an unknown) against an equal or better standard.\u00a0 A standard in a&#8230;","protected":false},"author":5,"featured_media":5023,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"inline_featured_image":false,"footnotes":""},"categories":[73],"tags":[],"class_list":["post-4884","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-calibration-guides"],"acf":[],"yoast_head":"<!-- This site is optimized with the Yoast SEO plugin v25.5 - https:\/\/yoast.com\/wordpress\/plugins\/seo\/ -->\n<title>Understanding As-Found and As-Left Data When It Comes to Calibration - SRP control systems ltd<\/title>\n<meta name=\"robots\" content=\"index, follow, max-snippet:-1, max-image-preview:large, max-video-preview:-1\" \/>\n<link rel=\"canonical\" href=\"https:\/\/www.srpcontrol.com\/understanding-as-found-and-as-left-data-when-it-comes-to-calibration\/\" \/>\n<meta property=\"og:locale\" content=\"en_US\" \/>\n<meta property=\"og:type\" content=\"article\" \/>\n<meta property=\"og:title\" content=\"Understanding As-Found and As-Left Data When It Comes to Calibration - SRP control systems ltd\" \/>\n<meta property=\"og:description\" content=\"Have you sent your instrument out for recalibration, gotten the certificate back, and seen &#8220;As-Found Calibration&#8221; or &#8220;As-Left Calibration&#8221; and wondered what this means? Here is a guide to understanding what these terms mean. The Definition of Calibration Calibration compares a measurement device (an unknown) against an equal or better standard.\u00a0 A standard in a...\" \/>\n<meta property=\"og:url\" content=\"https:\/\/www.srpcontrol.com\/understanding-as-found-and-as-left-data-when-it-comes-to-calibration\/\" \/>\n<meta property=\"og:site_name\" content=\"SRP control systems ltd\" \/>\n<meta property=\"article:published_time\" content=\"2020-04-28T14:53:15+00:00\" \/>\n<meta property=\"article:modified_time\" content=\"2023-09-21T18:44:59+00:00\" \/>\n<meta property=\"og:image\" content=\"https:\/\/www.srpcontrol.com\/wp-content\/uploads\/2023\/07\/IMG_E2012-scaled-e1689785754323.jpg\" \/>\n\t<meta property=\"og:image:width\" content=\"2560\" \/>\n\t<meta property=\"og:image:height\" content=\"1363\" \/>\n\t<meta property=\"og:image:type\" content=\"image\/jpeg\" \/>\n<meta name=\"author\" content=\"sara\" \/>\n<meta name=\"twitter:card\" content=\"summary_large_image\" \/>\n<meta name=\"twitter:label1\" content=\"Written by\" \/>\n\t<meta name=\"twitter:data1\" content=\"sara\" \/>\n\t<meta name=\"twitter:label2\" content=\"Est. reading time\" \/>\n\t<meta name=\"twitter:data2\" content=\"3 minutes\" \/>\n<script type=\"application\/ld+json\" class=\"yoast-schema-graph\">{\"@context\":\"https:\/\/schema.org\",\"@graph\":[{\"@type\":\"WebPage\",\"@id\":\"https:\/\/www.srpcontrol.com\/understanding-as-found-and-as-left-data-when-it-comes-to-calibration\/\",\"url\":\"https:\/\/www.srpcontrol.com\/understanding-as-found-and-as-left-data-when-it-comes-to-calibration\/\",\"name\":\"Understanding As-Found and As-Left Data When It Comes to Calibration - SRP control systems ltd\",\"isPartOf\":{\"@id\":\"https:\/\/www.srpcontrol.com\/#website\"},\"primaryImageOfPage\":{\"@id\":\"https:\/\/www.srpcontrol.com\/understanding-as-found-and-as-left-data-when-it-comes-to-calibration\/#primaryimage\"},\"image\":{\"@id\":\"https:\/\/www.srpcontrol.com\/understanding-as-found-and-as-left-data-when-it-comes-to-calibration\/#primaryimage\"},\"thumbnailUrl\":\"https:\/\/www.srpcontrol.com\/wp-content\/uploads\/2023\/07\/IMG_E2012-scaled-e1689785754323.jpg\",\"datePublished\":\"2020-04-28T14:53:15+00:00\",\"dateModified\":\"2023-09-21T18:44:59+00:00\",\"author\":{\"@id\":\"https:\/\/www.srpcontrol.com\/#\/schema\/person\/7fddec6920195fc39a98aed8805e981c\"},\"breadcrumb\":{\"@id\":\"https:\/\/www.srpcontrol.com\/understanding-as-found-and-as-left-data-when-it-comes-to-calibration\/#breadcrumb\"},\"inLanguage\":\"en-US\",\"potentialAction\":[{\"@type\":\"ReadAction\",\"target\":[\"https:\/\/www.srpcontrol.com\/understanding-as-found-and-as-left-data-when-it-comes-to-calibration\/\"]}]},{\"@type\":\"ImageObject\",\"inLanguage\":\"en-US\",\"@id\":\"https:\/\/www.srpcontrol.com\/understanding-as-found-and-as-left-data-when-it-comes-to-calibration\/#primaryimage\",\"url\":\"https:\/\/www.srpcontrol.com\/wp-content\/uploads\/2023\/07\/IMG_E2012-scaled-e1689785754323.jpg\",\"contentUrl\":\"https:\/\/www.srpcontrol.com\/wp-content\/uploads\/2023\/07\/IMG_E2012-scaled-e1689785754323.jpg\",\"width\":2560,\"height\":1363},{\"@type\":\"BreadcrumbList\",\"@id\":\"https:\/\/www.srpcontrol.com\/understanding-as-found-and-as-left-data-when-it-comes-to-calibration\/#breadcrumb\",\"itemListElement\":[{\"@type\":\"ListItem\",\"position\":1,\"name\":\"Home\",\"item\":\"https:\/\/www.srpcontrol.com\/\"},{\"@type\":\"ListItem\",\"position\":2,\"name\":\"Understanding As-Found and As-Left Data When It Comes to Calibration\"}]},{\"@type\":\"WebSite\",\"@id\":\"https:\/\/www.srpcontrol.com\/#website\",\"url\":\"https:\/\/www.srpcontrol.com\/\",\"name\":\"SRP control systems ltd\",\"description\":\"Instrumentation &amp; Calibration Instruments since 1980\",\"potentialAction\":[{\"@type\":\"SearchAction\",\"target\":{\"@type\":\"EntryPoint\",\"urlTemplate\":\"https:\/\/www.srpcontrol.com\/?s={search_term_string}\"},\"query-input\":{\"@type\":\"PropertyValueSpecification\",\"valueRequired\":true,\"valueName\":\"search_term_string\"}}],\"inLanguage\":\"en-US\"},{\"@type\":\"Person\",\"@id\":\"https:\/\/www.srpcontrol.com\/#\/schema\/person\/7fddec6920195fc39a98aed8805e981c\",\"name\":\"sara\",\"image\":{\"@type\":\"ImageObject\",\"inLanguage\":\"en-US\",\"@id\":\"https:\/\/www.srpcontrol.com\/#\/schema\/person\/image\/\",\"url\":\"https:\/\/secure.gravatar.com\/avatar\/ba65fba1080c39a4c0001777539727241f8d9d7cf9e191ab3cb67fa4a1a5ccb5?s=96&d=mm&r=g\",\"contentUrl\":\"https:\/\/secure.gravatar.com\/avatar\/ba65fba1080c39a4c0001777539727241f8d9d7cf9e191ab3cb67fa4a1a5ccb5?s=96&d=mm&r=g\",\"caption\":\"sara\"}}]}<\/script>\n<!-- \/ Yoast SEO plugin. -->","yoast_head_json":{"title":"Understanding As-Found and As-Left Data When It Comes to Calibration - SRP control systems ltd","robots":{"index":"index","follow":"follow","max-snippet":"max-snippet:-1","max-image-preview":"max-image-preview:large","max-video-preview":"max-video-preview:-1"},"canonical":"https:\/\/www.srpcontrol.com\/understanding-as-found-and-as-left-data-when-it-comes-to-calibration\/","og_locale":"en_US","og_type":"article","og_title":"Understanding As-Found and As-Left Data When It Comes to Calibration - SRP control systems ltd","og_description":"Have you sent your instrument out for recalibration, gotten the certificate back, and seen &#8220;As-Found Calibration&#8221; or &#8220;As-Left Calibration&#8221; and wondered what this means? Here is a guide to understanding what these terms mean. The Definition of Calibration Calibration compares a measurement device (an unknown) against an equal or better standard.\u00a0 A standard in a...","og_url":"https:\/\/www.srpcontrol.com\/understanding-as-found-and-as-left-data-when-it-comes-to-calibration\/","og_site_name":"SRP control systems ltd","article_published_time":"2020-04-28T14:53:15+00:00","article_modified_time":"2023-09-21T18:44:59+00:00","og_image":[{"width":2560,"height":1363,"url":"https:\/\/www.srpcontrol.com\/wp-content\/uploads\/2023\/07\/IMG_E2012-scaled-e1689785754323.jpg","type":"image\/jpeg"}],"author":"sara","twitter_card":"summary_large_image","twitter_misc":{"Written by":"sara","Est. reading time":"3 minutes"},"schema":{"@context":"https:\/\/schema.org","@graph":[{"@type":"WebPage","@id":"https:\/\/www.srpcontrol.com\/understanding-as-found-and-as-left-data-when-it-comes-to-calibration\/","url":"https:\/\/www.srpcontrol.com\/understanding-as-found-and-as-left-data-when-it-comes-to-calibration\/","name":"Understanding As-Found and As-Left Data When It Comes to Calibration - SRP control systems ltd","isPartOf":{"@id":"https:\/\/www.srpcontrol.com\/#website"},"primaryImageOfPage":{"@id":"https:\/\/www.srpcontrol.com\/understanding-as-found-and-as-left-data-when-it-comes-to-calibration\/#primaryimage"},"image":{"@id":"https:\/\/www.srpcontrol.com\/understanding-as-found-and-as-left-data-when-it-comes-to-calibration\/#primaryimage"},"thumbnailUrl":"https:\/\/www.srpcontrol.com\/wp-content\/uploads\/2023\/07\/IMG_E2012-scaled-e1689785754323.jpg","datePublished":"2020-04-28T14:53:15+00:00","dateModified":"2023-09-21T18:44:59+00:00","author":{"@id":"https:\/\/www.srpcontrol.com\/#\/schema\/person\/7fddec6920195fc39a98aed8805e981c"},"breadcrumb":{"@id":"https:\/\/www.srpcontrol.com\/understanding-as-found-and-as-left-data-when-it-comes-to-calibration\/#breadcrumb"},"inLanguage":"en-US","potentialAction":[{"@type":"ReadAction","target":["https:\/\/www.srpcontrol.com\/understanding-as-found-and-as-left-data-when-it-comes-to-calibration\/"]}]},{"@type":"ImageObject","inLanguage":"en-US","@id":"https:\/\/www.srpcontrol.com\/understanding-as-found-and-as-left-data-when-it-comes-to-calibration\/#primaryimage","url":"https:\/\/www.srpcontrol.com\/wp-content\/uploads\/2023\/07\/IMG_E2012-scaled-e1689785754323.jpg","contentUrl":"https:\/\/www.srpcontrol.com\/wp-content\/uploads\/2023\/07\/IMG_E2012-scaled-e1689785754323.jpg","width":2560,"height":1363},{"@type":"BreadcrumbList","@id":"https:\/\/www.srpcontrol.com\/understanding-as-found-and-as-left-data-when-it-comes-to-calibration\/#breadcrumb","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https:\/\/www.srpcontrol.com\/"},{"@type":"ListItem","position":2,"name":"Understanding As-Found and As-Left Data When It Comes to Calibration"}]},{"@type":"WebSite","@id":"https:\/\/www.srpcontrol.com\/#website","url":"https:\/\/www.srpcontrol.com\/","name":"SRP control systems ltd","description":"Instrumentation &amp; Calibration Instruments since 1980","potentialAction":[{"@type":"SearchAction","target":{"@type":"EntryPoint","urlTemplate":"https:\/\/www.srpcontrol.com\/?s={search_term_string}"},"query-input":{"@type":"PropertyValueSpecification","valueRequired":true,"valueName":"search_term_string"}}],"inLanguage":"en-US"},{"@type":"Person","@id":"https:\/\/www.srpcontrol.com\/#\/schema\/person\/7fddec6920195fc39a98aed8805e981c","name":"sara","image":{"@type":"ImageObject","inLanguage":"en-US","@id":"https:\/\/www.srpcontrol.com\/#\/schema\/person\/image\/","url":"https:\/\/secure.gravatar.com\/avatar\/ba65fba1080c39a4c0001777539727241f8d9d7cf9e191ab3cb67fa4a1a5ccb5?s=96&d=mm&r=g","contentUrl":"https:\/\/secure.gravatar.com\/avatar\/ba65fba1080c39a4c0001777539727241f8d9d7cf9e191ab3cb67fa4a1a5ccb5?s=96&d=mm&r=g","caption":"sara"}}]}},"_links":{"self":[{"href":"https:\/\/www.srpcontrol.com\/wp-json\/wp\/v2\/posts\/4884","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.srpcontrol.com\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.srpcontrol.com\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.srpcontrol.com\/wp-json\/wp\/v2\/users\/5"}],"replies":[{"embeddable":true,"href":"https:\/\/www.srpcontrol.com\/wp-json\/wp\/v2\/comments?post=4884"}],"version-history":[{"count":1,"href":"https:\/\/www.srpcontrol.com\/wp-json\/wp\/v2\/posts\/4884\/revisions"}],"predecessor-version":[{"id":4885,"href":"https:\/\/www.srpcontrol.com\/wp-json\/wp\/v2\/posts\/4884\/revisions\/4885"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.srpcontrol.com\/wp-json\/wp\/v2\/media\/5023"}],"wp:attachment":[{"href":"https:\/\/www.srpcontrol.com\/wp-json\/wp\/v2\/media?parent=4884"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.srpcontrol.com\/wp-json\/wp\/v2\/categories?post=4884"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.srpcontrol.com\/wp-json\/wp\/v2\/tags?post=4884"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}